- Measurement item
-
Total Organic Carbon (TOC)
-
- Measurement method
-
UV + Sodium persulfate oxidization method / NDIR detection method
-
- Measurement range
-
0~30mg/L(Other ranges are available upon request)
-
- Measurement interval
-
10min (Adjustable)
-
- Detection limit
-
0.05mg/L F.S.
-
- Accuracy
-
±2% F.S.
-
- Reproducibility
-
±2% F.S.
-
- Linearity
-
±3% F.S.
-
- Zero/Span drift
-
±2% F.S.
-
- Calibration
-
Auto/manual calibration by standard solution
-
- Sample/Reagent measurement method
-
Pulse control, Reagent loop injection system
-
- Sample/Reagent distribution method
-
6-port distribution control method (Fixed quantity injection with multi-channel valve system)
-
-
-
- Operation system
-
Windows 7 (CPU process) or more
-
- Output
-
TCP/IP, DC 4~20mA, RS-232C
-
- Status display
-
Normal operation, on calibration, error
-
- Display
-
Color TFT LCD 8″ (Touch screen)
-
- Power
-
AC 220V, 50/60Hz
-
- Power consumption
-
Average 70W
-
- Voltage regulation
-
3% or less
-
- Data storage capacity
-
HDD 100G or more
-
- Data output
-
USB terminal
-
- Enclosure / Size
-
Steel、IP55 / 500(W) ⅹ 335(D) ⅹ 750(H) mm
-
- Operating humidity
-
RH 95% or less
-
SẢN PHẨM
HATA-4000 Online TOC Analyzer
HATA-4000 Online TOC Analyzer
UV Persulphuric Acid Oxidation-type Online Total Organic Carbon (TOC) Analyzer, HATA-4000, is a measuring equipment that has automated the “Continuous Automatic TOC Measuring Method” of the Standard Water Pollutant Testing Method. HATA-4000 has minimized quantitative errors of sample analysis by using peristaltic metering pump and sample loop system, and it is measuring equipment that has been developed to accurately and easily measure TOC by improving the precision through the sensor measurement value correction function according to the gas flow rate. Through the application of UV persulphuric acid oxidation method, it has wide range of applications from rivers to wastewater.